Adjustments / Programs 5 - 17
Siemens AG SPR2-130.061.02 Page 17 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Slot diaphragm 5
Prerequisites 5
• The setting of the camera rotation is correctly adjusted and is in the 0° position.
• Connect the service PC to the serial interface of the SIREMOBIL.
• Start the service program.
• Select the "Adjustments, Calibrations..." menu. Answer the query with "Yes".
• Select "SCREEN DISPLAY SLOT DIAPHRAGM" in the "Function Groups:" combo box.
Disable circle mask
• Switch off zoom.
• In the "Available Functions:" window, select "DISABLE CIRCLE MASK" and click on "Ex-
ecute". Close the status window by clicking on the OK button.
For the query "Do you want to reset the Unit", click on "No".
Start adjustment
• In the "Available Functions:" window select "INIT" and click on "Execute". Close the sta-
tus window by clicking on the OK button.
For the query "Do you want to reset the Unit", click on "No".
Full format, slot diaphragm open
• Move the camera into the 0 degree position.
The X-Iris opens to survey format. To open it further, use the keys for opening the X-Iris.
The slot diaphragm is opened automatically.
• Apply fluoro briefly.
• Open and close the slot diaphragm display via the keys.
• Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
• In the window "Available Functions:" select "Slot Diaphragm Open Pos - I.I. Full Format"
and click "Execute".
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No". The slot dia-
phragm is closed automatically.
Full format, slot diaphragm closed
• Apply fluoro briefly.
• Open and close the slot diaphragm display via the keys.
• Align the lines shown on the monitor with the front edges of the slot diaphragm blades.
• In the window "Available Functions:" select "Slot Diaphragm Close Pos - I.I. Full Format"
and click "Execute".
• Close the status window by clicking on the OK button.
Use the lowest KV values possible to ensure optimal detection of
the front edges of the slot diaphragm blades.
NOTE