7 - 1
Siemens AG SPR2-130.061.02 Page 1 of 2 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Brake force / Brake torque 7
Brake forces 7
• Position prerequisites for measuring the forces: angulation, orbital, and horizontal swivel
movements must be at zero on the scale and horizontal movement should be at 20 on
the scale.
• All measurements should be made with a spring measure on the railing near the image
intensifier with the DHHS spacer, if present, installed on the X-ray tube.
SIREMOBIL Compact L 7
SIREMOBIL Compact 7
Risk of physical injury and damage to property!
When performing service work on the horizontal slide of the basic
unit (for example when performing repair work on the brake of the
C-arm horizontal travel) it is possible that the C-arm may slide out
of its position and drop onto the floor after the end stop has been
removed.
Before dismantling the end stop, the horizontal travel brake must
be locked and the horizontal slide must be mechanically locked in
order to prevent it from sliding out. For example, the basic unit
can be positioned in the room in such a way that the C-arm
mechanically touches one of the walls.
See also page 1- 8 "Back stop plate at the C-arm horizontal
support"
Movement Nominal values Comments
Brake released: Brake
applied:
Horizontal travel 10 N to 40 N 160 N to 200 N n.a.
Angulation movement 10 Nm to 20 Nm 60 Nm to 80 Nm n.a.
Orbital movement 10 Nm to 24 Nm 60 Nm to 80 Nm measure with the C-arm
in vertical position.
Horizontal swivel 5 Nm to 10 Nm 90 Nm to 150 Nm n.a.
Movement Nominal values Comments
Brake released: Brake applied:
Horizontal travel >
20 N / < 35 N > 160 N n.a.
Angulation <
17 Nm > 40 Nm n.a.
Orbital movement >
10 N / < 24 N > 65 N measure with the C-arm in
vertical position.
Horizontal swivel <
10 Nm > 90 Nm n.a.
WARNING