Replacing boards / replacing components 6 - 9
Siemens AG SPR2-130.061.02 Page 9 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
• The C-arm can now be rotated to permit easier installation of components (I.I. above,
tube assembly underneath).
• Reinstall the O-ring and the cap nut on the anode cable.
- The cable length from the beginning of the white wire to the cap nut should be 70 mm.
• Reinstall the high voltage section in the C-arm.
- Reconnect all the cables.
• Reinstall the hand rail above the VIDEOMED DC.
• Reinstall both jacks D4.X1 for the VIDEOMED DC and X45 for the cassette contact on
the receptacle.
• Reconnect the D4.X1 plug for the VIDEOMED DC and the X45 plug for the cassette con-
tact.
• Reconnect plug M30.X1.
• Reconnect plug D120.X1 to the control section of the I.I. mini power supply.
• Reconnect the protective conductor to the ground screw (5/Fig. 1).
• Reinstall the cable clips (4/Fig. 1).
• Reinstall the counterbalance weight for the 7"/17 cm I.I.
Check the existence of the sealing ring 6 /Fig. 3.
Checks and adjustments 6
• Check the I.I. electrode voltages according to the test protocol for the I.I.
-Do not readjust in cases where there are only slight deviations from the values on the
test protocol (measurement device tolerances).
• Check that the camera optics are centered to the I.I. output and adjust them if neces-
sary; refer to "Replacing the camera optics," subitem "Checking the centering of the
camera optics to the I.I. output " and "Centering the camera optics to the I.I. output".
• Check the camera rotation and readjust it if necessary.
• Check the setting of the X-iris and readjust it if necessary.
• Check the display of the blades on the monitor and readjust if necessary.
• Check the dose rate and readjust it if necessary.
• Perform the IQ Quick test.
- When doing this, check the resolution first and, if necessary, readjust the optical reso-
lution of the Compact optics.
• Complete the country-specific acceptance (§16 Partial acceptance... /DHHS...).
Check the presence of the sealing ring 6/Fig. 3 at the new I.I.
If none is present, remove the sealing ring from the old I.I. (glued
to the I.I. housing with 4 to 6 dots), lay it in the groove of the new
I.I. housing and fix it with 4 to 6 dots of adhesive
(e.g. superglue).
NOTE