EasyManua.ls Logo

Siemens SIREMOBIL Compact L - Checks and adjustments

Siemens SIREMOBIL Compact L
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacing boards / replacing components 6 - 9
Siemens AG SPR2-130.061.02 Page 9 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
The C-arm can now be rotated to permit easier installation of components (I.I. above,
tube assembly underneath).
Reinstall the O-ring and the cap nut on the anode cable.
- The cable length from the beginning of the white wire to the cap nut should be 70 mm.
Reinstall the high voltage section in the C-arm.
- Reconnect all the cables.
Reinstall the hand rail above the VIDEOMED DC.
Reinstall both jacks D4.X1 for the VIDEOMED DC and X45 for the cassette contact on
the receptacle.
Reconnect the D4.X1 plug for the VIDEOMED DC and the X45 plug for the cassette con-
tact.
Reconnect plug M30.X1.
Reconnect plug D120.X1 to the control section of the I.I. mini power supply.
Reconnect the protective conductor to the ground screw (5/Fig. 1).
Reinstall the cable clips (4/Fig. 1).
Reinstall the counterbalance weight for the 7"/17 cm I.I.
Check the existence of the sealing ring 6 /Fig. 3.
Checks and adjustments 6
Check the I.I. electrode voltages according to the test protocol for the I.I.
-Do not readjust in cases where there are only slight deviations from the values on the
test protocol (measurement device tolerances).
Check that the camera optics are centered to the I.I. output and adjust them if neces-
sary; refer to "Replacing the camera optics," subitem "Checking the centering of the
camera optics to the I.I. output " and "Centering the camera optics to the I.I. output".
Check the camera rotation and readjust it if necessary.
Check the setting of the X-iris and readjust it if necessary.
Check the display of the blades on the monitor and readjust if necessary.
Check the dose rate and readjust it if necessary.
Perform the IQ Quick test.
- When doing this, check the resolution first and, if necessary, readjust the optical reso-
lution of the Compact optics.
Complete the country-specific acceptance (§16 Partial acceptance... /DHHS...).
Check the presence of the sealing ring 6/Fig. 3 at the new I.I.
If none is present, remove the sealing ring from the old I.I. (glued
to the I.I. housing with 4 to 6 dots), lay it in the groove of the new
I.I. housing and fix it with 4 to 6 dots of adhesive
(e.g. superglue).
NOTE

Table of Contents

Other manuals for Siemens SIREMOBIL Compact L

Related product manuals