Replacing boards / replacing components 6 - 13
Siemens AG SPR2-130.061.02 Page 13 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Rotate the camera in the in +220° position.
Mark the center again on the monitor (see Fig. 9).
Place an 8 mm washer over the three markings.
All three markings must be located within the 8 mm diameter of the washer. An 8 mm
washer has a somewhat larger inner diameter of 8.4 mm, therefore the markings must
clearly lie within the inner diameter and should not touch the edge of the washer.
If the diameter is <= 8 mm, the following centering adjustment does not have to be per-
formed.
Continue with "checks and adjustments".
If the markings exceed > 8 mm, proceed as follows:
Centering the camera optics to the I.I. output 6
Prerequisite: The +220°, -220° and 0° positions were already marked as previously
described.
Rotate the camera optics back to the 0° position.
Slightly loosen the three attachment screws of the camera optics.
Slightly loosen both excenter screws adjacent to the camera optics.
Move the camera optics toward the center, between both +220° and -220° markings (see
also Fig. 9).
Switch fluoro on briefly and evaluate the position on the monitor. According to the ratio of
the I.I. input screen diameter to the I.I. output screen diameter, a shift of the camera op-
tics by approximately 0.1 mm corresponds to a position shift of the washer displayed on
the monitor of more than 1.3 mm (17 cm I.I.) or 1.7 mm (23 cm I.I.).
Fig. 9
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING