Replacing boards / replacing components 6 - 23
Siemens AG SPR2-130.061.02 Page 23 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
MEMOSKOP C-SUB/ MEMOSKOP C-SUB & MOD 6
Replace the MEMOSKOP C-SUB.
The attachment screws are accessible on the underside of the storage device.
Check the software status of the MEMOSKOP. Download the system software status to
the MEMOSKOP if necessary.
Use the MEMOSKOP Service Software to import the language file.
Set the MEMOSKOP C-SUB to the correct video frequency and configure it for the num-
ber of existing monitors.
Refer to Chapter 5 "Configuring the video frequency " and "MEMOSKOP - Configuration
for the number of existing monitors".
Configure the following parameters in Technical Setup for MEMOSKOP:
Check the software status of the newly delivered MEMOSKOP.
If the software status of the MEMOSKOP differs from that of the
system, the system software status must be downloaded to the
MEMOSKOP. The MEMOSKOP software status is displayed in the
2nd line of the User Setup. The first number describes the
MEMOSKOP type (2=CE; 3=CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB
& MOD).
Read and record the parameters for MEMOSKOP C-SUB if you
can select User setup.
After replacing, configure the MEMOSKOP C-SUB according to
the parameters recorded.
Number of monitors Configure 1 for SIREMOBIL Compact with 1 monitor.
Configure 2 for SIREMOBIL Compact with 2 monitors.
Hardcopy Check whether 5000 is programmed in "4 Max Hard-
copy time". If not, program this value.
Edge 1 coef set Check whether 3 is programmed, If not, program this
value.
Edge 2 coef set Check whether 8 is programmed, If not, program this
value.
Motion detection setup Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not,
program these values.
Date / Time format Set the country-specific format.
Keyboard Set to the keyboard layout being used.
Language Set for language of the country.
Display of realtime clock Configure Yes. Select No only upon customer request.
Display of rotated image Configure Yes.
Monitor zoom Configure No.