SIREMOBIL Compact SPR2-130.061.02 Page 24 of 28 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
6 - 24 Replacing boards / replacing components
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number Configure 800.
Disable dose rate selection
in user setup
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on Disk" and
menu "Dose on Hardcopy output" in User Setup are vis-
ible. Otherwise, set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL Compact.
Do not enter the first zero.
Subtraction K-Factor,
phase A
Program "32". Program a different value only upon cus-
tomer request.
Subtraction K-Factor,
phase B
Program "MD 2". Program a different value only upon
customer request.
Product name SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default Program changeable Default: Yes.
Depending on country specific regulations, configure
Yes or No.
Create External disk If a MOD-drive is present or the option "DICOM bridge"
is installed, configure Yes. Without MOD drive or
DICOM bridge, configure No.
Enable Bridge If the option "DICOM bridge" is installed, configure Yes.
Otherwise configure No.
Software backward compatible Configure No.
Default LUT Configure 2. Only upon customers request program a
different LUT curve.
Default Edge Configure 1. Only upon customers request program a
different. Edge enhancement value
Default LUT for subtraction Configure 1. Only upon customers request program a
different LUT curve.
Subtraction Time of phase A Program "4.5".
Program a different value only upon customer request.
Roadmap K-Factor, phase A Program "32".
Program a different value only upon customer request.
Roadmap K-Factor, phase B Program "MD 2".
Program a different value only upon customer request.
Roadmap K-Factor, phase C Program "MD 2".
Program a different value only upon customer request.
Roadmap Time of phase A Program "4.5".
Program a different value only upon customer request.