Adjustments / Programs 5 - 13
Siemens AG SPR2-130.061.02 Page 13 of 30 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
• Click on "OK" . The new position values for the collimator will be stored on the disk.
• Return to the main menu with the ESC key.
Collimator, slot diaphragm 5
Prerequisites 5
• Camera rotation has been adjusted correctly.
• Image reversal functions are disabled.
• The "SLOT DIAPHRAGM," "TYPE OF DIAPHRAGM" parameter has been programmed
for "DISPLAY YES".
Adjustment 5
• Connect the service PC to the serial service interface.
• Start the service program.
• Select the Adjustment, Calibrations...menu.
• Respond to the query with yes.
• In the Combo box "Function Groups:" select "Slot Diaphragm".
• In the window "Available Functions:" select "Disable circle mask" and click "Execute.
The circle mask of the Memoskop is switched off.
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No" .
• In the window "Available Functions:" select "Init" and the "Execute".
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No" .
• The mechanical limits are checked and the position values are automatically stored.
• Camera rotation is automatically moved to the 0 degree position.
• The X-Iris is opened to max.aperture.
• Switch off Zoom .
• Attach a centering cross or equivalent object to the I.I. for evaluating the position of the
fluoro image
• Switch on fluoro
• Move the slot diaphragm to the 0 degree position using the keys for rotating the slot dia-
phragm. The blades should be in the vertical on the monitor screen.
• Switch off fluoro.
• In the window "Available Functions:" select "0 DEGREE POSITION" and click "Execute".
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING