Replacing boards / replacing components 6 - 5
Siemens AG SPR2-130.061.02 Page 5 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
SIREPHOS 6
Open the SIREPHOS cover (see Chapter 1, "Covers").
Remove the laser targeting device, if present.
Remove the dose measurement chamber, if present.
Remove the collimator.
Disconnect the SIREPHOS plug.
Remove the ground screw for the protective conductor.
Remove the rubber stop of the C-arm on the SIREPHOS side. The screws that secure
the SIREPHOS are now visible.
Remove the SIREPHOS screws. When doing this, have a second person hold the X-ray
tube unit.
Place the new X-ray tube unit on the guide bolts and secure it with both Allen screws.
Ensure that all cables are routed properly.
Reconnect the SIREPHOS plug and secure it.
Reconnect the protective conductor using the ground screw.
Do not install the rubber stop yet.
Reinstall the collimator and connect it.
Reinstall the dose measurement chamber / laser targeting device and connect them.
Perform the kV offset adjustment.
Perform the generator adjustment (filament circuit learning program).
Ensure that the collimator is functioning properly and is set correctly.
Ensure that the collimator blades are centered in the blanking circle.
- If possible, slightly loosen the Sirephos screws and tilt the SIREPHOS to adjust it.
- Retighten the SIREPHOS screws.
Ensure that the collimator blades of the iris diaphragm are visible during fluoroscopy in
the survey format and zoom format on at least two sides of the monitor image.
Check the display of the blades on the monitor screen and adjust if necessary.
Check collimation for direct exposure if there is a cassette holder.
Reinstall the rubber stop.
Ensure that the area/dose product measurement device is functioning properly
(if present).
Check the dose rate settings and adjust them if necessary.
Check the settings of the maximum skin dose rate.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
WARNING