SIREMOBIL Compact SPR2-130.061.02 Page 14 of 30 Siemens AG
System Manual Rev. 11 09.04 CS PS 24 Medical Solutions
5 - 14 Adjustments / Programs
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No" .
• Switch on fluoro.
• Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
• Switch off fluoro.
• In the window "Available Functions:" select "FULL FORMAT POSITION" and click "Exe-
cute".
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No".
• Select the zoom format.
• Switch on fluoro.
• Open the slot diaphragm to where the blades begin to disappear at the edge of the im-
age.
• Switch off fluoro.
• In the window "Available Functions:" select "ZOOM FORMAT POSITION" and click "Ex-
ecute".
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "No" .
• In the window "Available Functions:" select "Enable circle mask" and click "Execute"
• Respond to the query "Do you want to reset the unit" by clicking "No" .
• In the window "Available Functions:" select "INIT OFF & STORE POSITION" and click
"Execute".
• Close the status window by clicking on the OK button.
• Respond to the query "Do you want to reset the unit" by clicking "Yes". The adjustment
values are stored.
Concluding steps 5
• Switch off the system and switch it back on again.
• Return to the main menu by pressing ESC at the service PC.
• Select the "Data, Backup..." menu.
• Click "parameter" in the Backup window.
• Then click "Backup". Enter "slot diaphragm adjusted" under "Remarks for Parameter".
• Then click "OK" . The new position values are stored to disk.
• You can now exit from the service program.