EasyManua.ls Logo

Siemens SIREMOBIL Compact L - Page 77

Siemens SIREMOBIL Compact L
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Replacing boards / replacing components 6 - 25
Siemens AG SPR2-130.061.02 Page 25 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
Configure the following parameters in User setup for MEMOSKOP:
Perform the IQ quick test.
Operating mode Parameters Change the programmed parameters only upon cus-
tomer request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Split screen Configure horizontally.
Configure vertically only upon customer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.
Hospital Name Program the hospital.
Storage of Dose on Disk If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.
Dose on hardcopy output If area dose product meter is present, (Diamentor) con-
figure Yes. Otherwise configure No.

Table of Contents

Other manuals for Siemens SIREMOBIL Compact L

Related product manuals