Replacing boards / replacing components 6 - 21
Siemens AG SPR2-130.061.02 Page 21 of 28 SIREMOBIL Compact
Medical Solutions Rev. 11 09.04 CS PS 24 System Manual
• Configure the following parameters in User setup for MEMOSKOP:
Monitor text Configure Yes.
Maximum image number Configure 700.
Disable dose rate selection
in user setup
Set to "No" if equipped with an area dose product mea-
surement device. Menu "Storage of Dose on Disk" and
menu "Dose on Hardcopy output" in User Setup are vis-
ible. Otherwise, set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL Compact,
Do not enter the first zero.
Default LUT Configure 2. Only upon customers request program a
different LUT curve.
Default Edge Configure 1. Only upon customers request program a
different. Edge enhancement value
Software backward compatible Configure No.
Product name SIREMOBIL Compact (default). If wrong, configure
"SIREMOBIL Compact".
Default Program changeable Default: Yes.
Depending on country specific regulations, configure
Yes or No.
Create External disk If a MOD-drive is present or the option "DICOM bridge"
is installed, configure Yes. Without MOD drive or
DICOM bridge, configure No.
Enable Bridge If the option "DICOM bridge" is installed, configure Yes.
Otherwise configure No.
Operating mode Parameters Change the programmed parameters only upon cus-
tomer request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
Text monitor For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Split screen Configure horizontally. Configure vertically only upon
customer request.
Display of time Configure Yes.
Set Date Time Set the current date and time.