Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
1010 Float / 4 Path 2 offset User calibrated zero off‐
set for path 2.
Calibrated delta time =
delta time + Path 2 fac‐
tory offset + Path 2 off‐
set.
0.0
[s]
-1.25E-6 - +1.25E-6 Read /
write
1014 Float / 4 Path 3 offset User calibrated zero off‐
set for path 3.
Calibrated delta time =
delta time + Path 3 fac‐
tory offset + Path 3 off‐
set.
0.0
[s]
-1.25E-6 - +1.25E-6 Read /
write
1018 Float / 4 Path 4 offset User calibrated zero off‐
set for path 4.
Calibrated delta time =
delta time + Path 4 fac‐
tory offset + Path 4 off‐
set.
0.0
[s]
-1.25E-6 - +1.25E-6 Read /
write
1022 Float / 4 Path 5 offset User calibrated zero off‐
set for path 5.
Calibrated delta time =
delta time + Path 5 fac‐
tory offset + Path 5 off‐
set.
0.0
[s]
-1.25E-6 - +1.25E-6 Read /
write
1004 Float / 4 Path 1 factory
offset
Factory calibrated zero
offset for path 1.
0.0
[s]
- Read
only
1008 Float / 4 Path 2 factory
offset
Factory calibrated zero
offset for path 2.
0.0
[s]
- Read
only
1012 Float / 4 Path 3 factory
offset
Factory calibrated zero
offset for path 3.
0.0
[s]
- Read
only
1016 Float / 4 Path 4 factory
offset
Factory calibrated zero
offset for path 4.
0.0
[s]
- Read
only
1020 Float / 4 Path 5 factory
offset
Factory calibrated zero
offset for path 5.
0.0
[s]
- Read
only
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 173