Service and maintenance
6
6.1 Basic safety notes
Note
The device is maintenance-free.
6.1.1 Maintenance
The device is maintenance-free. However, a periodic inspection according to pertinent directives
and regulations must be carried out.
An inspection can include:
Ambient conditions
Seal integrity of the process connections, cable entries, and cover
Reliability of power supply, lightning protection, and grounds
WARNING
Dust layers above 5 mm
Risk of explosion in hazardous areas.
Device may overheat due to dust build up.
Remove dust layers in excess of 5mm.
WARNING
Use of a computer in a hazardous area
If the interface to the computer is used in the hazardous area, there is a risk of explosion.
Ensure that the atmosphere is explosion-free (hot work permit).
NOTICE
Penetration of moisture into the device
Damage to device.
Make sure when carrying out cleaning and maintenance work that no moisture penetrates
the inside of the device.