Selectable Analog current
(neg. MV on)
Analog current
(neg. MV o󹪜)
4 - 20 mA 4 mA 4 mA
NAMUR - 20 mA 4 mA 4 mA
Table 8-3 Full-scale value of the analog output current for upward limiting of measured value
Selectable Analog current
(neg. MV on)
Analog current
(neg. MV o󹪜)
0 - 20 mA 21.0 mA 21.0 mA
2 - 20 mA 21.0 mA
1)
21.0 mA
1)
4 - 20 mA 21.0 mA
1)
21.0 mA
1)
NAMUR - 20 mA 20.5 mA 20.5 mA
1)
If measuring range 2 is set to the maximum possible value, the limiting value of range 2 ... 20 mA is at
20.9 mA and of range 4 ... 20 mA at 20.8 mA.
Analog output during FCTRL (analog output during function control)
Note
If a fault and a function control are present on the analyzer at the same time, only the values of
the setting "Output during faults" apply. Simultaneous occurrence of values of the setting
"Output during FCTRL" are ignored in this case.
Analog output during
function control
Meas. value: actual
84112
The status "FCTRL" (function control) is set
During an AUTOCAL procedure
During the warm-up phase
During a calibration procedure
During remote control via the communication interface
In the uncoded state:
Output of the following measured values is possible here:
Hold: The value measured directly prior to commencement of a function control is output
unchanged. This also applies to the limits which are output (see section Parameters: Limits
(Page 130)).
Actual: The measured value is continuously updated.
Functions
8.4 Con󹪝guration
ULTRAMAT 23
138 Equipment Manual, 04/2022, A5E37100388-012