EasyManua.ls Logo

Siemens ULTRAMAT 23 Series - Page 139

Siemens ULTRAMAT 23 Series
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Zero:
See following table:
Selectable Analog current
0 - 20 mA 0 mA
2 - 20 mA 2 mA
4 - 20 mA 4 mA
NAMUR - 20 mA 3 mA
21 mA:
see following table:
Selectable Analog current
0 - 20 mA 21.0 mA
2 - 20 mA 21.0 mA
4 - 20 mA 21.0 mA
NAMUR - 20 mA 21.5 mA
Analog output during fault
Note
If a fault and a function control are present on the analyzer at the same time, only the values of
the setting "Output during faults" apply. Simultaneous occurrence of values of the setting
"Output during FCTRL" are ignored in this case.
Analog output
during faults
Meas. value: ZERO
84113
Here, you can de󹪝ne the type of measured value output during
a fault.
Output of the following measured values is possible here:
Hold: The value measured directly prior to commencement of a fault is output unchanged.
This also applies to the limits which are output (see section Parameters: Limits (Page 130)).
Actual: The measured value is continuously updated.
Functions
8.4 Con󹪝guration
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 139

Table of Contents

Related product manuals