• Version with IR detector with chopper compartment purging:
This version consumes approx. 100 ml/min of purging gas; you must set an inlet pressure of
300 kPa (3 bar).
• Version with UV absorption photometer for small SO
2
measuring ranges and NO
2
measurements
Benets
• AUTOCAL can be carried out with ambient air (zero-point calibration for the IR and UV
components and full-scale calibration as well as calibration point for oxygen sensors) and is
therefore highly cost eective as no calibration gases and accessories are required
• Monitoring of the ow rate using a pressure switch, alarm when the ow rate falls below the
nominal ow rate
• High selectivity thanks to multi-layer IR detectors,
low cross-sensitivity to water vapor
• Sample chambers can be cleaned (for analysis chamber lengths of 20mm and more),
resulting in cost savings through reuse following contamination
• Menu-assisted operation in plain text, thus high operational safety
• Service information and logbook, cost savings through preventive maintenance and help for
service and maintenance personnel
• Increased safety through coded operator levels, thus protection against unauthorized access
or clumsy working
• Open interface architecture (ELAN (RS485), PROFIBUS-DP/PA), thus simplied process
integration
• Communication software. SIMATIC PDM and SIPROM GA.
• Remote operation and control (via SIPROM GA and PROFIBUS).
Special benets when used in biogas plants
• Continuous measurement of all 4 important components, including H
2
S
• Long service life of the H
2
S sensor even at increased concentrations;
no diluting or backushing necessary
Description
3.1 Area of application
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012 21