3.7 Safety devices
Pos: 66 /100 Sig ma/100 BA Zentrif ugen Sigm a (Standard module)/030 Sic herheit/030-0070-0 010 Deckel verriegelung @ 27\mod_140 5319036321_68 .docx @ 192171 @ 3 @ 1
3.7.1 Lid lock device
The centrifuge can only be started when the lid is properly closed. The
electrical lock must be locked. The lid can only be opened when the rotor
has stopped. If the lid is opened by way of the emergency release system
during operation, the centrifuge will immediately switch off and decelerate
brakeless. If the lid is open, the drive is completely separated from the
mains power supply, i.e. the centrifuge cannot be started (see chapter 7.1.1
- "Emergency lid release").
Pos: 67 /010 Uni versalmodule/ Leerz eile 8pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 68 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0070-0 020 Stillstan dsüberwachu ng @ 27\mod_1 405319037803 _68.docx @ 192185 @ 3 @ 1
3.7.2 Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill.
This standstill is checked by the microprocessor.
Pos: 69 /010 Uni versalmodule/ Leerz eile 8pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 70 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0070-0 030 Syste mkontrolle Spinco ntrol S @ 27\ mod_14053190 39294_68.doc x @ 192199 @ 3 @ 1
3.7.3 System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. Errors are detected with extreme sensitivity and
displayed as error messages in a dialog box (see chapter 7.2 - "Table of
error codes").
Pos: 71 /010 Uni versalmodule/ Leerzeile 8 pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 72 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sic herheit/030-0070-0 040 Schutzleit erprüfung mit P otentialausg leichsschraube @ 27\mod_1 405319043129_ 68.docx @ 1922 41 @ 3 @ 1
3.7.4 Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on
the rear panel of the centrifuge (see chapter 2.1.1 - "Functional and
operating elements"). An earth conductor check can be carried out by
authorized and specialized personnel using a suitable measuring
instrument. Please contact the service department (see chapter 7.3 -
"Service contact").
Pos: 73 /010 Uni versalmodule/ Leerz eile 8pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 74 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0070-0 050 Unw uchtüberwachung Spincontrol S @ 27\mod_140 5319045827_ 68.docx @ 192 269 @ 3 @ 1
3.7.5 Imbalance monitoring system
A dialog box may pop up or emit a sound signal in order to indicate that the
centrifuge is in the inadmissible imbalance range. If the rotor is loaded
unevenly, the drive will be switched off in the acceleration phase or during
the run.
Pos: 75 /010 Uni versalmodule/ Leerzeile 8 pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 76 /100 Sig ma/100 BA Zentrif ugen Sigm a (Standard module)/030 Sic herheit/030-0070-0 060 Tempera turüberwach ung @ 27\mod _1405319048 438_68.docx @ 19 2297 @ 3 @ 1
3.7.6 Temperature monitoring system
If the temperature inside the rotor chamber rises above 50°C, the drive
system will be switched off automatically. The centrifuge cannot be
restarted until it has cooled.
Pos: 77 /010 Uni versalmodule/ Leerzeil e 8pt @ 12\mo d_133706306 2114_0.docx @ 65687 @ @ 1
Pos: 78 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/030 Sicher heit/030-0070-0 070 Rotorü berwachung @ 2 7\mod_140 5319049932_68.d ocx @ 1923 11 @ 3 @ 1
3.7.7 Rotor monitoring system
When a rotor number and, if applicable, a bucket number are selected, the
computer will automatically check whether the entered speed or the
entered gravitational field are permissible for the selected rotor.