Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
8 Maintenance and service
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
75 / 131
Translation of the original operating manual, part no. 0702502
Pos: 205 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0010-0 010 Kondensator @ 28\mod_1 405320616708_6 8.docx @ 19 4831 @ 4 @ 1
8.1.1.1 Condenser (only refrigerated centrifuges with an air-cooled refrigeration system)
In order to cool the refrigerant that is compressed by the refrigeration unit,
centrifuges with an air-cooled refrigeration system use a lamellar
condenser. It is cooled by air.
Dust and dirt obstruct the cooling flow of air. The dust on condenser pipes
and lamellas reduces the heat exchange and thus the performance of the
refrigeration unit.
This is why the installation site should be as clean as possible.
Check the condenser at least once a month for dirt and clean it if
necessary.
If you have any queries, please contact service (see chapter 7.3 -
"Service contact").
Pos: 206 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 207 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8K S_8KBS_8KHB S/080 Wartung und Instandhalt ung/080-001 0-0020 Zube hör @ 29\ mod_140532747205 7_68.docx @ 1 99913 @ 3 @ 1
8.1.2 Accessories
CAUTION
For the care of the accessories, special safety measures must be
considered as these are measures that will ensure operational safety at
the same time!
WARNING
All swing-out rotors applicable for this centrifuge and the angle rotor 12510
weigh more than 18 kg.
Always lift the rotors with a lifting device or with a sufficient number of
people helping you.
Immediately rinse off the rotor, buckets, or accessories under running
water if they have come into contact with any liquids that may cause
corrosion. Use a brush for test tubes to clean the bores of angle rotors.
Turn the rotor upside down and allow it to dry completely.
Clean the accessories outside the centrifuge once a week or preferably
after each use. Adapters should be removed, cleaned and dried.
CAUTION
Do not clean the accessories in a dishwasher!
Cleaning in a dishwasher removes the anodised coating; the result is
cracking in areas that are subject to stress.
If the rotors or accessories have been contaminated with toxic,
radioactive, or pathogenic substances, clean them immediately with a
suitable decontamination agent (depending on the type of
contamination). Take suitable precautions for your own safety if there is
a risk of toxic, radioactive, or pathogenic contamination.
Dry the accessories with a soft cloth or in a drying chamber at
approx. 50°C.