Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
8 Maintenance and service
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
77 / 131
Translation of the original operating manual, part no. 0702502
Pos: 213 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0040 Trag bolzen (außer 2-6, 2-6E, 2-7 , 2-16P, 2-1 6KL) @ 28\mod_ 14053206247 92_68.docx @ 19 4915 @ 3 @ 1
8.1.4 Load bearing bolts
Only greased load-bearing bolts ensure a uniform swing-out of the buckets
and, therefore, the smooth operation of the centrifuge. Load-bearing bolts
that are insufficiently greased may cause the centrifuge to stop due to an
imbalance.
Clean the load-bearing bolts and bucket groove in order to remove the
old grease.
Apply a small amount of heavy-duty grease for load-bearing bolts (ref.
no. 71401, see the following picture) to both load-bearing bolts of a
bucket.
Fig. 45: Sufficient quantity of grease for one bolt
Install the bucket and swing it manually back and forth once in order to
distribute the grease.
Repeat this process with all the other buckets.
Pos: 214 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312 _0.docx @ 105 @ @ 1