Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
8 Maintenance and service
Version 08/2012, Rev. 2.5 of 02/07/2018 • sb
Translation of the original operating manual, part no. 0702502
Pos: 221 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartu ng und Instandh altung/080-0 030 Instan dhaltungsarbeit en @ 28\mod_14 05320628810 _68.docx @ 19 4957 @ 2 @ 1
8.3 Service
In the event of service work that requires the removal of the panels, there
is a risk of electric shock or mechanical injury. Only qualified specialist
personnel is authorised to perform this service work.
The centrifuge is subject to high mechanical stress. In order to be able to
withstand this high level of stress, high-quality components were used
during the production of the centrifuge. Nevertheless, wear cannot be
excluded and it may not be visible from the outside. Especially the rubber
parts that are – among other things – part of the motor suspension, are
subject to ageing.
This is why we recommend having the centrifuge checked by the
manufacturer during an inspection once per year in the operating state and
once every three years in the dismantled state. Rubber parts should be
replaced after three years.
Information and appointments:
In Germany:
Contact
Sigma Laborzentrifugen GmbH
An der Unteren Söse 50
37520 Osterode (Germany)
Tel. +49 (0) 55 22 / 50 07-44 44
E-mail: support.lab@sigma-zentrifugen.de
Outside Germany:
Contact our agency in your country. All agencies are listed at
www.sigma-zentrifugen.de [Sales Partners]
• If you would like to utilise our service, please state the type of your
centrifuge and its serial number.
Pos: 222 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312 _0.docx @ 105 @ @ 1