Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
7 Malfunctions and error correction
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
71 / 131
Translation of the original operating manual, part no. 0702502
Pos: 193 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8K S_8KBS_8KHB S/070 Störunge n und Fehlers uche/070-0 010-0011 Not entriegelung des D eckels - flac her Deckel @ 46\mod_1481 537260198_68.doc x @ 461692 @ 3 @ 1
7.1.1 Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge
lid.
Switch off the mains power switch and disconnect the power cord from
the socket.
Remove the plugs (see figure, item 1) at the right side panel, e.g. with a
screw driver.
Fig. 43: Position of the openings for the emergency lid release
Insert the supplied tube wrench (part no. 930 110) horizontally into the
hole. The key will be guided through a funnel-shaped tube to the shaft
of the lid lock motor.
Fig. 44: The emergency lid release key must be inserted horizontally.
Unlock the motorised lid locks as follows:
Turn the left lid lock anti-clockwise.
Turn the right lid lock clockwise.
Then, reinsert the plugs.
WARNING
The lid may only be unlocked and opened when the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge
run, the centrifuge will be switched off immediately and decelerate in an
unbraked manner.