Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
8 Maintenance and service
Version 08/2012, Rev. 2.5 of 02/07/2018 • sb
Translation of the original operating manual, part no. 0702502
Pos: 209 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0020-0 010 Kunststoffz ubehör @ 28\ mod_1405320 619602_68.docx @ 194859 @ 4 @ 1
8.1.2.1 Plastic accessories
The chemical resistance of plastic decreases with rising temperatures (see
chapter 11.5 - "Resistance data").
• If solvents, acids, or alkaline solutions have been used, clean the plastic
accessories thoroughly.
Plastic accessories must not be greased!
Pos: 210 /010 Uni versalmodule/ Leerz eile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 211 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/080 Wartu ng und Instandh altung/080-0 010-0030 Rot or, Becher u nd Vielfachträger ( außer Basic) @ 28\mod_14 05320621063 _68.docx @ 19487 3 @ 3 @ 1
8.1.3 Rotors, buckets and carriers
Rotors, buckets and carriers are produced with the highest precision, in
order to withstand the permanent high stress from high gravitational fields.
Chemical reactions as well as stress-corrosion (combination of oscillating
pressure and chemical reaction) can affect or destroy the metals. Barely
detectable cracks on the surface can expand and weaken the material
without any visible signs.
• Check the material regularly (at least once a month) for
– cracks
– visible damage of the surface
– pressure marks
– signs of corrosion
– other changes.
• Check the bores of the rotors and multiple carriers.
• Replace any damaged components immediately for your own safety.
• After every cleaning process, grease the rotor tie-down screw slightly
with a small amount of heavy-duty grease for load-bearing bolts (part
no. 71401) and distribute the grease with a cloth so that it forms a thin
layer.
Pos: 212 /010 Uni versalmodul e/Seitenwechsel @ 0\mod_12 02116244312 _0.docx @ 105 @ @ 1