5.6 Special equipment: water cooling system
Centrifuges with water cooling are equipped with a special refrigeration unit
with a refrigerant (CFC-free), modified electronic system, and a special
software version. The housing of the GMP version is completely closed. All
of the other units have vent slots for the motor and electronic power
system.
Operating conditions
The centrifuge must only be operated with media-neutral water of
drinking water quality (hardness <8) or with another suitable refrigerant.
We recommend using a filter (to be provided by the customer).
The operating pressure of the water must be between 1.5 and 5 bar.
We recommend using a pressure reducer (to be provided by the
customer).
The maximum flow rate depends on the operating conditions (e.g.
speed and rotor temperature). The minimum cooling water consumption
at maximum power is approximately 5 l/min.
At a standstill, the water supply is stopped by a solenoid valve. In the
case of the GMP version, the valve is opened in two steps:
1. reduced flow, only for the internal air cooler
2. unlimited flow, when the compressors are running.
The maximum temperature at the water inlet is 20°C. The lower the
water temperature, the more efficient the cooling.
The centrifuge is equipped with a ¾-inch water connector .
1 Connector for the
cooling water inlet
2 Connector for the
cooling water outlet
Fig. 18: Connections for the water cooling system
Pos: 119 /010 Uni versalmodule/ Abschnitts wechsel @ 0\ mod_1202124514 062_0.docx @ 418 @ @ 1