5 Set-up and connection
Pos: 97 /100 Sig ma/110 BA Zentrif ugen Sigma (Pr ojekte)/8KS_ 8KBS_8KHBS /050 Aufstellung und Anschluss/0 50-0010 Aufstellort 8KS _8KBS @ 29\ mod_1405327 400003_68.docx @ 199717 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to ambient air.
• Air cooled centrifuges: Ensure sufficient ventilation.
The centrifuge 8KS / SKBS with an air-cooled compressor should not be
set up with its left side against a wall, since otherwise the hot air, which is
emitted out the back, will be drawn in again as fresh air for cooling. As a
result, the unit will switch off due to overheating (see chapter 7.2 - "Table
of error codes").
• Water cooled centrifuges: Provide sufficient water throughput.
• Do not position the centrifuge near heat generators.
• Avoid direct sunlight (UV radiation).
• During transport from cold to warmer places, condensational water will
collect inside the centrifuge. Allow sufficient time for drying (min. 24 h)
before using the centrifuge again.
Pos: 98 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 99 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 100 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8KS _8KBS_8KHBS/ 050 Aufstellung und Anschluss/ 050-001 1 Aufstellen und Ausrichten d er Zentrifuge @ 2 9\mod_14053 27400697_68. docx @ 19973 1 @ 2 @ 1
5.2 Set-up and alignment of the centrifuge
Ensure that the centrifuge is disconnected from the power supply during
the set-up and alignment.
Pay attention to the layout plan (see chapter 11.9 - "Layout plan")!