5 Set-up and connection
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
33 / 131
Translation of the original operating manual, part no. 0702502
Pos: 96 /100 Sig ma/100 BA Zentrif ugen Sigma (S tandardmod ule)/050 Aufstellu ng und Anschl uss/050 Aufst ellung und A nschluss===== ======= ========== ============ @ 27\ mod_1405319 173071_68.doc x @ 192381 @ 1 @ 1
5 Set-up and connection
Pos: 97 /100 Sig ma/110 BA Zentrif ugen Sigma (Pr ojekte)/8KS_ 8KBS_8KHBS /050 Aufstellung und Anschluss/0 50-0010 Aufstellort 8KS _8KBS @ 29\ mod_1405327 400003_68.docx @ 199717 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to ambient air.
Air cooled centrifuges: Ensure sufficient ventilation.
NOTE
The centrifuge 8KS / SKBS with an air-cooled compressor should not be
set up with its left side against a wall, since otherwise the hot air, which is
emitted out the back, will be drawn in again as fresh air for cooling. As a
result, the unit will switch off due to overheating (see chapter 7.2 - "Table
of error codes").
Water cooled centrifuges: Provide sufficient water throughput.
Do not position the centrifuge near heat generators.
Avoid direct sunlight (UV radiation).
During transport from cold to warmer places, condensational water will
collect inside the centrifuge. Allow sufficient time for drying (min. 24 h)
before using the centrifuge again.
Pos: 98 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 99 /010 Uni versalmodule/ Leerzeile @ 0\mod_120 2116244500_ 0.docx @ 11 4 @ @ 1
Pos: 100 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8KS _8KBS_8KHBS/ 050 Aufstellung und Anschluss/ 050-001 1 Aufstellen und Ausrichten d er Zentrifuge @ 2 9\mod_14053 27400697_68. docx @ 19973 1 @ 2 @ 1
5.2 Set-up and alignment of the centrifuge
DANGER
Ensure that the centrifuge is disconnected from the power supply during
the set-up and alignment.
NOTE
Pay attention to the layout plan (see chapter 11.9 - "Layout plan")!
Pos: 101 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312 _0.docx @ 105 @ @ 1