EasyManua.ls Logo

Sigma 8KBS - Parameters Menu

Sigma 8KBS
131 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
6 Using the centrifuge
54 / 131
Pos: 159 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/060 Betrieb _Spincontr ol L+S/Spinco ntrol S (Zent7)/0 60-0030-0020-0 060 Menü P arameter - nur KS @ 27\mod_1 405320218640 _68.docx @ 193305 @ 4 @ 1
6.3.2.7 Parameters menu
The "Parameters" menu is symbolised by the " " symbol on the menu
bar. It is used to specify various conditions for the centrifugation. These
conditions are used to monitor the process and to control access to the
centrifuge.
Fig. 30: Menu "Parameters"
Process
Standstill cooling
Depending on the substances to be centrifuged, it may make sense to
precool the centrifuge. The precooling prevents the cooled samples in the
uncooled centrifuge from heating up to an inadmissible temperature level.
If the standstill cooling function is activated, the centrifuge starts to precool
after it is switched on. In the runtime field, the symbol "" is displayed. The
lid must be closed.
CAUTION
Unmoved air in the rotor chamber leads to an incorrect measuring and
control behaviour and causes the compressor to freeze over. At
temperatures below 0°C, aqueous liquids will freeze, thereby preventing
sedimentation.
Do not cool the rotor below 0°C at a standstill.
NOTE
The centrifuge is equipped with the "RAPID_TEMP" program. This
program is used to precool the rotor chamber quickly under defined
conditions (see chapter 6.3.2.5 - "Standard menu", Program list).
Radius
The radius determines the value of the relative centrifugal force (RCF) that
the sample is subjected to. Normally, the maximum RCF value is displayed.
If the value is reduced manually, a downward facing arrow "" will be
displayed in the RCF field.

Table of Contents

Related product manuals