6 Using the centrifuge
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
41 / 131
Translation of the original operating manual, part no. 0702502
Pos: 129 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8KS _8KBS_8KHBS/ 060 Betrieb/06 0-0020-002 0 Einsetzen von Rotoren und Zubehör 8KS_ 8KBS_8KHBS @ 29\mod_ 1405327449337 _68.docx @ 199 843 @ 3 @ 1
6.2.2 Installation of rotors and accessories
WARNING
All swing-out rotors applicable for this centrifuge and the angle rotor 12510
weigh more than 18 kg.
Always lift the rotors with a lifting device or with a sufficient number of
people helping you.
Pos: 130 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 131 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8K S_8KBS_8KHB S/060 Betrieb/ 060-0020-002 0-0010 Eins etzen eines Rot ors @ 29\mod_ 14053274500 12_68.docx @ 1 99857 @ 4 @ 1
6.2.2.1 Installation of a rotor
Open the centrifuge lid by pressing the lid key.
Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
Lower the rotor with its central bore straight down onto the motor shaft.
Turn the adapter end of the motor shaft anti-clockwise with the supplied
hexagon socket key (part no. 930 050) and tighten the internal thread of
the rotor with 20 Nm with the rotor wrench (size 17/19, part no. 930 018).
WARNING
After frequent use, the rotor tie-down screw must be loosened by some
turns, the rotor has to be lifted and fastened again. This must be done
once a day or after 20 cycles. This ensures a proper connection between
the rotor and the motor shaft.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
Pos: 132 /010 Uni versalmodule/ Leerzeile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 133 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8KS _8KBS_8KHBS/ 060 Betrieb/ 060-0020-002 0-0020 Eins etzen eines Winkelro tors mit Herm etikdeckel @ 29\mod_140 5327450800_ 68.docx @ 199871 @ 4 @ 1
6.2.2.2 Installation of an angle rotor with a hermetic lid
Lower the rotor onto the motor shaft.
Turn the adapter end of the motor shaft anti-clockwise with the supplied
hexagon socket key (part no. 930 050) and tighten the internal thread of
the rotor with 20 Nm with the rotor wrench (size 17/19, part no. 930 018).
NOTE
Slightly grease the rotor and lid seals after cleaning.
Screw the rotor cover onto the rotor and tighten it with the supplied tool
no. 17985 by hand.
The rotors can be installed or removed with a closed lid after loosening
the rotor tie-down screw.
Follow the safety instructions and hazard warnings (see chapter 3 -
"Safety")!
WARNING
The lid screw serves for the fastening of the lid onto the rotor only, not for
the fastening of the rotor onto the motor shaft!