Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
7 Malfunctions and error correction
Version 08/2012, Rev. 2.5 of 02/07/2018 • sb
Translation of the original operating manual, part no. 0702502
Pos: 195 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/070 Störu ngen und Fehlersu che/070-00 20 Tabelle der Fehlermeld ungen @ 28\ mod_1405320582 996_68.doc x @ 194761 @ 2 @ 1
7.2 Table of error codes
• Allow to slow down
• Power off/on
All these errors stop the
centrifuge or cause it to
decelerate brakeless
• Allow to slow down
• Power off/on
• Power off
• Ensure ventilation
• Allow to slow down
• Power off/on
With error 34, 35, and 36, the
centrifuge will stop;
with error 37 and 38 only an
error message will be given
Temperature error
(only for refrigerated centrifuges)
• Allow to slow down
• Power off
• Allow to cool down
• Provide better ventilation (only
air cooled centrifuges)
• Provide sufficient water
throughput (only water cooled
centrifuges)
Imbalance error (only for centrifuges
with imbalance monitoring system)
• Allow to slow down
• Power off
• Eliminate the imbalance
• Press lid key
• Close lid
• Remove foreign matter from
the opening of the lid lock
device
With error 50 and 51, the
centrifuge will stop
• Allow to slow down
• Power off/on
With error 60, the message
"power failure during run"will
be displayed, with error 61,
the message "stop after
power on" will be displayed
• Allow to slow down
• Power off/on
• Power off
• Allow to cool down
• Provide for better ventilation
With error 83, error message
only
• Check connections
• Provide sufficient water
throughput (only water cooled
centrifuges)
If it is impossible to eliminate the errors, contact the service!
Pos: 196 /010 Uni versalmodule/ Seitenwechsel @ 0\mod_12 02116244312 _0.docx @ 105 @ @ 1