Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
5 Set-up and connection
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
37 / 131
Translation of the original operating manual, part no. 0702502
Pos: 109 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/050 Aufst ellung und Anschlu ss/050-00 20 Energie versorgung------------------------------------------------------ @ 27\mod_14053191756 30_68.docx @ 192409 @ 2 @ 1
5.4 Power supply
Pos: 110 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/050 Aufst ellung und Anschlu ss/050-00 20-0013 Ansc hlussart (Cekon- Stecker 5-polig) @ 32\mod_ 14175998430 77_68.docx @ 2 22470 @ 3 @ 1
5.4.1 Type of connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
SIGMA laboratory centrifuges are units of protection class I. The
centrifuges of this type have a five-wire power cord (2.5 m) with a 5-
channel Cekon-plug (16 A). Behind the front door there are three fuses with
a rocker switch.
Pos: 111 /010 Uni versalmodule/ Leerz eile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 112 /100 Sig ma/100 BA Ze ntrifugen Sig ma (Standardmo dule)/050 Aufst ellung und Anschlu ss/050-00 20-0020 Sic herungen bausei ts (außer 3-3 0KS) @ 7\mod_1 30631230239 7_68.docx @ 4 3513 @ 3 @ 1
5.4.2 Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that
are to be provided by the customer.
Pos: 113 /010 Uni versalmodule/ Leerz eile @ 0\mod_12 02116244500 _0.docx @ 1 14 @ @ 1
Pos: 114 /100 Sig ma/110 BA Ze ntrifugen Sig ma (Projekte)/8KS _8KBS_8KHBS/ 050 Aufstellung und Anschluss/ 050-004 0-0030 Trenn vorrichtung @ 29\mod_140532 7407717_68. docx @ 199801 @ 3 @ 1
5.4.3 Power isolating device
The electrical installation of the building must include a power isolating
device in the form of a switch or circuit breaker. This device must be
located near the centrifuge. Furthermore, it must be easily accessible for
the operator and marked as the power isolating device for the centrifuge.