Refrigerated Centrifuge Sigma 8KS
Refrigerated Centrifuge Sigma 8KBS
6 Using the centrifuge
Version 08/2012, Rev. 2.5 of 02/07/2018 sb
57 / 131
Translation of the original operating manual, part no. 0702502
Lock
In order to prevent any unauthorised use of the centrifuge, the following
functions can be blocked:
Saving of programs (level 01)
Changing of parameters, short run (level 02)
Loading of programs (level 03)
Start button (level 04)
Blocking a function
Select the function that is to be blocked. The lower levels will also be
automatically selected (if, for example, the "Parameters" function is
selected, the "Save" function will also be selected).
Select the button "Activate code".
Enter a four-digit code and confirm the entry.
The lock is now active. The symbol " " will be displayed in the status line
and the lock level will be indicated.
If changes are made after a function has been blocked, the system will ask
for the code prior to executing the change.
Fig. 32: "Parameters" menu with an active lock (level 01)
Unblocking a function
Select the button "Deactivate code".
Enter the code and confirm the entry.
The lock is now deactivated.
Changing the code
Select the button "Change code".
Enter the old code and confirm the entry.
Enter the new code.
For safety reasons, the code must be entered a second time.
The code is now changed.