Technical Information
67
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
The pump is intended for use in the electromagnetic environment specied below. The customer or the user of the pump should assure that it is used
in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be used no closer to any part of
the pump, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Conducted RF,
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not
applicable
Recommended separation distance
d=[3,5/V1]*P
½
Radiated RF,
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
15 V/m
d=0.23*P
½
80MHz to 800MHz
d=0.47*P
½
800MHz to 2,5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as determined by an electromagnetic site survey
a
,
should be less than the compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures,
objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF
transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the pump is used exceeds
the applicable RF compliance level above, the pump should be observed to verify normal operation. If abnormal performance is observed, addi-
tional measures may be necessary, such as reorienting or relocating the pump.
b
(Not applicable) Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than [V₁] V/m.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals