Technical Information
66
Guidance and Manufacturer’s Declaration Electromagnetic Immunity
The pump is intended for use in the electromagnetic environment specied below. The customer or the user of the pump should assure that it is
used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Not applicable
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
IEC 61000-4-5
± 1 kV dierential mode
± 2 kV common mode
Not applicable
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interruptions
and voltage variations on power
supply input lines, IEC 61000-4-11
< 5 % UT (> 95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
< 5 % UT (> 95 % dip in UT)
for 5 sec
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the Pump requires
continued operation during power mains interruptions, it is
recommended that the Pump be powered from an uninter-
ruptible power supply or a battery.
Power frequency (50/60 Hz)
Magnetic eld IEC 61000-4-8
3 A/m 400 A/m
(IEC 60601-2-24)
Power frequency magnetic elds should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Smiths Medical CADD-MS 3 and is the answer not in the manual?

Smiths Medical CADD-MS 3 Specifications

General IconGeneral
BrandSmiths Medical
ModelCADD-MS 3
CategoryMedical Equipment
LanguageEnglish

Related product manuals