Chapter 4: Measurements and Calculations 81
Meas. & Calcs
OB Calculations
References for measurements and calculations are in Chapter 8, “References.” The following table
shows the transducers and exam types that provide OB calculations.
View or Repeat
Saved
Measurement
1 Highlight the desired measurement from the calculations menu.
The saved measurement is shown at the bottom of the calculation
menu.
2 Press the Select key or the Caliper key to select the measurement.
3 Repeat the measurement.
The new results are displayed in the measurement and calculation
data area. You can compare the active measurement to the saved
measurement.
4 To save the new measurement, select Save from the on-screen menu,
or press the Enter key.
Note: This will save the new measurement to the patient report and
overwrite the previously saved measurement.
Delete
Measurement
1 Highlight the measurement.
2 Select Delete from the on-screen menu.
This will remove the last saved measurement from the patient report.
Warning: Make sure you have selected the OB exam type and the OB calcs author for the OB
table you intend to use. See Table 9, “System Defined OB Calculations and Table
Authors” on page 83.
In OB exams, an accurate date and time are critical for accurate obstetrics
calculations. Verify that the date and time are accurate before each use of the
system. The system does not automatically adjust for daylight savings time
changes.
Verify the patient information, date, and time settings are accurate.
Before starting a new calculation, start a new patient exam to delete the previous
measurements. See “New Patient” on page 41.
Prior to use, verify OB custom table data entries are correct. The system does not
confirm the accuracy of the custom table data entered by the user.
Table 7: Transducers and Exam Type
Transducer Exam Type
C8 OB
C15 OB
C60 OB
ICT OB