Command Descriptions
VX4792 User Manual
3-23
CH1
<Auto Step Data Block>
<Space>
?
DEFine
:AUTOSTep
<Space> <File Name>
,
CH1
,
,
<File Name>
<File Name>::=<string>
which is the name of the file to which the auto step data is transmitted.
CH1
This argument is included for compatibility purposes only.
<Auto Step Data Block>::=<Arbitrary Data Block>
where <Arbitrary Data Block> consists of auto step data written in ASCII
code, with the data of each step separated with a Line Feed (LF) code. The
following diagram and examples reflect these requirements.
#232WAVE01.WFM<LF>WAVE02.WFM<LF>WAVE03.WFM
 
  
   
  
  
AUTOSTEP:DEFINE "AUTOS01.AST",
#232WAVE01.WFM<LF>WAVE02.WFM<LF>WAVE03.WFM
sends the auto step data to the file AUTOS01.AST
*CAL?
The *CAL? common query performs an internal calibration and returns status that
indicates whether the waveform generator completes the self-calibration without
error. If an error is detected during calibration, execution immediately stops and
an error code is returned.
. The waveform generator does not respond to any commands or queries
issued during the self-cal. The self-cal takes up to 30 seconds to complete.
CALIBRATION and DIAGNOSTIC
Arguments
Examples
Group