EasyManua.ls Logo

Sony Tektronix VX4792 - Page 260

Sony Tektronix VX4792
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Verification
VX4792 User Manual
F-27
This procedure checks the CLOCK OUTPUT signal amplitude.
Electrical Characteristic Checked: Auxiliary Outputs, Clock, Amplitude, on
page E–6.
Equipment Required: A 50 coaxial cable and an oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install test setup and set test equipment controls (see Figure F–13):


 


Figure F-13: Clock Amplitude Initial Test Setup
a. Connect oscilloscope: Connect the CLOCK OUTPUT through a coaxial
cable to the oscilloscope CH1 vertical input.
b. Set oscilloscope controls:
Vertical CH1
Coupling DC
Scale 200 mV/div
Input impedance 50
Horizontal
Sweep 500 ns/div
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500 mV
Mode Auto
Check Clock Amplitude

Table of Contents