Step Size 7,2mV 3,1mV 1,27mV 318µV
Noise (with filterbox) < 850µV < 400µV < 100µV < 50µV
3,5kV Bi Analy :
Range +/- 3500V +/- 1500V
Step Size 7,1mV 1,7mV
Noise (with filterbox) < 600µV < 500µV
1.4 Safety Instructions
Before performing any electrical or electronic operations, please consult ‘SPECS
Safety Instructions’ and follow them closely. Please consult as well any documenta-
tion of supplied material from third party manufacturers before any operation and
strictly follow the safety instructions given therein.
1.5 Initial Inspection and Packaging List
Visual inspection of the component is required to ensure that no damage has oc-
curred during shipping. Should there be any signs of damage or parts missing, con-
tact SPECS immediately. Please check the delivery according to the packaging list for
completeness.
Quantity Item
1 HSA3500 or HSA3500 plus
1 Printed HSA3500 Manual
1 Power cord (200V or 115V)
2 CAN-Bus cable
2 CAN-Bus terminator
2 HV cable (Channel HV, Channel Base)
1 HV cable (Screen) optional
1 HV Interlock override connector
1 EC10 Can-Ethernet-Adapter
1 Ethernet cable
HSA3500 / HSA3500 plus Power Supply 3