10V: 50V,60V,...,200V. If there are any other DAC voltages to
change at the same time, these voltages will also be changed in
15 equal steps, even if their MaxStep limit is deactivated.
Default: 0.0 (deactivated if less or equal 0.0)
DelayUp unsigned
long,
optional
If this entry exists and is greater than 0, it gives the minimal
delay between two DAC output operations for increasing output
voltages
(|U1| < |U2|).
default: 0 (ignored if 0)
Units: ms
DelayDown unsigned
long,
optional
If this entry exists and is greater than 0, it gives the minimal
delay between two DAC output operations for decreasing output
voltages
(|U2| < |U1|).
Default: 0 (ignored if 0)
Units: ms
StandByValue string Value of DAC Voltages while the Unit is in standby. Can be given
as an absolute (e.g. 1000) or relative value (e.g. 20%).
StandByValuePP string Same as above for positive polarity (detection of ions).
Table 25: ModeDef section description
An example for a [ModeDef] section is:
[ModeDef] LargeArea:3.5kV ###############################################
{
MinRR = 0.0
MaxRR = 1000.0
ShortName = LA:3.5kV
RelaisMode = (0,0) # no relays used
[Parameter] Ekin
{
Formula = $Ekin
Unit = eV
}
[Parameter] Epass
{
Formula = $Ep
Unit = eV
}
[Parameter] NonDisp
{
Formula = %1/10 + %2
Unit = mm
}
[LogicalVoltage] "Kinetic Energy"
HSA3500 / HSA3500 plus Power Supply 41