Technical data
SPECTRO XEPOS — 19.01.2017 — 17
3 Technical data
Description Value
Instrument height 465 mm (18.3“)
Instrument width 618 mm (24.3“)
Instrument depth 730 mm (28.8“)
Instrument weight 65 kg (143 lbs) approx.
Place of installation Inside of dry, dust-free rooms up to an
altitude of 2,000 m above sea level,
vibration-resistant (not subject to impact
and vibration loads)
Operating voltage range 90 - 253 VAC, 50/60 Hz
Spectrometer power input max. 1,200 W (without PC)
Fuse 16 A (slow-blow)
Temperature range
Storage room temperature range
+10 to +35°C (50 to 95°F)
-10 to +50°C (14 to 122°F)
Optimal temperature range (within
specifications)
+20°C – +25°C (68 °F – 77 °F)
Air humidity 10 to 80% (non-condensing) at 25°C
10 to 75% (non-condensing) at 30 °C
Protection class/contamination class IP20 (DIN 40050), contamination class II
(laboratory environment)
EMC environment Normal laboratory environment
Helium gas flushing (option)
Pressure 1.0 – 3.0 bar; (15 – 44 psi)
Quality ≥ 99.996% (4.6)
Consumption typically 4.5 L per analysis
Vacuum (option)
End pressure/operating pressure < 5 mbar (0.07 psi)
Voltage supply 110 V/230 V (instrument-controlled)