QA BEAMCHECKER™ PLUS 28
6.5 Deleting Baselines
If a plan or energy was created by mistake or is no longer needed, it can be deleted from the
database using the following procedure.
1. Connect the QA BeamChecker Plus to the PC.
2. Launch the Communication Software and ensure the correct database is loaded.
3. Select the desired room containing the plan or energy for deletion on the left side of the
screen.
4. Select the Baselines tab and confirm the Connection Status reads “Connected, Waiting for
User” (yellow).
5. Select the desired plan or energy for deletion.
6. Click the Delete Plan or Delete Energy button depending on the room type.
7. Confirm the operation.
CAUTION: Once an energy or plan is deleted, it CANNOT be recovered.