EasyManuals Logo

Standard Imaging QA BEAMCHECKER PLUS User Manual

Standard Imaging QA BEAMCHECKER PLUS
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
QA BEAMCHECKER™ PLUS 65
RTP_LTE_RAW
Data collected from the right (photon side) or left (electron side)
ionization chamber during measurement
BOTTOM_RAW
Data collected from the bottom ionization chamber during
measurement
ONE_MM_ALUM_RAW
Data collected from energy channel 1 during measurement, primarily
used for electrons
SIX_MM_ALUM_RAW
Data collected from energy channel 2 during measurement, primarily
used for electrons
SIX_MM_LEAD_RAW
Data collected from energy channel 3 during measurement, primarily
used for photons
NUM_SAMPLES
Number of samples collected by the analog-to-digital converter
during the measurement
COMMENT
Measurement comment entered using the Data View interface
14.2 TomoTherapy Mode - Static Plans
When exporting from the TomoTherapy Mode, an individual file is created for each plan. The
format of each file is broken down into multiple sections using the following pattern: Baseline1;
Measurements for Baseline1; Baseline2; Measurements for Baseline2; etc
Baseline Data
Data Label
Description
BASELINE_ID
A unique identifier for the baseline energy that incorporates date and
time of measurement
FIELD_SIZE
Collimator size, as entered in the Baseline Setup screen
SSD
Source to Surface Distance, as entered in the Baseline Setup screen
DOSERATE
Linear accelerator dose rate, as entered in the Baseline Setup screen
DOSE
Linear accelerator dose, as entered in the Baseline Setup screen
GANTRYANGLE
Gantry angle, as entered in the Baseline Setup screen
TEMPERATURE
Internal temperature when baseline measurement was taken, in °C
PRESSURE
Pressure when baseline measurement was taken, in mmHg
ENERGYLEVEL
A unique identifier for the baseline energy as determined by several
ionization chambers and a proprietary algorithm
LAT_PROFILE
Lateral profile of baseline, calculated from the ionization chamber
readings
---
---
OUTPUT_MEASURE
Output measurement of baseline, calculated from the ionization
chamber readings
---
---
ENERGY_MEASURE
Baseline energy measurement, calculated from the ionization
chamber readings
---
---

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Standard Imaging QA BEAMCHECKER PLUS and is the answer not in the manual?

Standard Imaging QA BEAMCHECKER PLUS Specifications

General IconGeneral
BrandStandard Imaging
ModelQA BEAMCHECKER PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals