EasyManuals Logo

Standard Imaging QA BEAMCHECKER PLUS User Manual

Standard Imaging QA BEAMCHECKER PLUS
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
QA BEAMCHECKER™ PLUS 69
LEFT_OUTPUT_%DELTA
Left channel output difference from baseline
CENTER_OUTPUT
Center channel output (temperature/pressure corrected)
CENTER_OUTPUT_%DELTA
Center channel output percentage difference from baseline
RIGHT_OUTPUT
Right channel output (temperature/pressure corrected)
RIGHT_OUTPUT_%DELTA
Right channel output percentage difference from baseline
CENTER_RAW
Data collected from the center ionization chamber during
measurement
LTP_RTE_RAW
Data collected from the left (photon side) or right (electron side)
ionization chamber during measurement
TOP_RAW
Data collected from the top ionization chamber during
measurement
RTP_LTE_RAW
Data collected from the right (photon side) or left (electron side)
ionization chamber during measurement
BOTTOM_RAW
Data collected from the bottom ionization chamber during
measurement
ONE_MM_ALUM_RAW
Data collected from energy channel 1 during measurement,
primarily used for electrons
SIX_MM_ALUM_RAW
Data collected from energy channel 2 during measurement,
primarily used for electrons
SIX_MM_LEAD_RAW
Data collected from energy channel 3 during measurement,
primarily used for photons
NUM_SAMPLES
Number of samples collected by the analog-to-digital converter
during the measurement
COMMENT
Measurement comment entered using the Data View interface
14.4 Dynamic 5 Channel Mode
When exporting from the Dynamic 5 Channel Mode, an individual file is created for each plan.
The format of each file is broken down into multiple sections using the following pattern:
Baseline1; Measurements for Baseline1; Baseline2; Measurements for Baseline2; etc
Baseline Data
Data Label
Description
BASELINE_ID
A unique identifier for the baseline energy that incorporates date and
time of measurement
FIELD_SIZE
Collimator size, as entered in the Baseline Setup screen
SSD
Source to Surface Distance, as entered in the Baseline Setup screen
DOSERATE
Linear accelerator dose rate, as entered in the Baseline Setup screen
DOSE
Linear accelerator dose, as entered in the Baseline Setup screen
GANTRYANGLE
Gantry angle, as entered in the Baseline Setup screen
TEMPERATURE
Internal temperature when baseline measurement was taken, in °C
PRESSURE
Pressure when baseline measurement was taken, in mmHg

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Standard Imaging QA BEAMCHECKER PLUS and is the answer not in the manual?

Standard Imaging QA BEAMCHECKER PLUS Specifications

General IconGeneral
BrandStandard Imaging
ModelQA BEAMCHECKER PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals