QA BEAMCHECKERâ„¢ PLUS 52
The left column shows the room, energy, and plan information that exists within the database,
and the right column shows which information is stored within the QA BeamChecker Plus
internal memory. By clicking either Yes or Synchronize, the values will be pushed from left to
right and the main QA BeamChecker Plus software interface will be displayed.
If the QA BeamChecker Plus was synchronized to the incorrect database, simply open the
correct database file and re-perform the synchronize operation.
NOTE: If using QA BeamChecker Plus software version 2.2.2 and above, Wire-Free
measurement data will not be deleted during the synchronization process. Upon the
completion of the synchronization operation, go to the Data View tab to download any Wire-
Free data to the currently opened database.
What to do if Synchronization is Not Allowed
In QA BeamChecker Plus software versions 2.2.1 and below, if Wire-Free mode measurement
data is present within the QA BeamChecker Plus internal memory, a synchronization
operation is not permitted. The data will first need to be downloaded to the correct database
file before proceeding. If the correct database file cannot be located, the measurement data
must be erased from the unit before proceeding. This can be accomplished by first creating a
new database, then re-synchronizing to the desired database.
If version 2.2.2 and above is used, synchronization IS permitted even if measurement data
remains on the unit. Once synchronization is performed, data will remain on the QA
BeamChecker Plus and can be downloaded to the database once synchronization is
completed.
CAUTION: Ensure you are using the correct database file before proceeding with the
download process. If data is transferred to the incorrect database, it cannot be moved to
another database without assistance from Standard Imaging.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Standard Imaging QA BEAMCHECKER PLUS and is the answer not in the manual?

Standard Imaging QA BEAMCHECKER PLUS Specifications

General IconGeneral
BrandStandard Imaging
ModelQA BEAMCHECKER PLUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals