Instruction Manual for Libra Care Bed
Page 102
6.5.4.1 Installing the battery
Proceed as follows to install the rechargeable battery on the cross tube on the mattress
base frame at the head end:
1. Insert four expanding nuts into the holes in the battery mounting plate A .
2. Place the battery mounting plate on the cross tube from above B .
3. Place the battery on the mounting plate from below and fasten it with four screws
(4 x 16 mm) C .
4. Connect the battery to the control unit (see chapter 6.5.4.2).
The battery cable must point towards the centre of the bed. The battery cable must be routed
and fixed in place with cable ties in such a way that it cannot be pulled off, driven over,
twisted, stretched or crushed when the bed is adjusted.