Instruction Manual for Libra Care Bed
Page 32
4.3.3 Trendelenburg handset (optional)
The Trendelenburg handset is optional and is used when using the bed for long-term
medical care.
4.3.3.1 Adjustment functions
Pressing this button will
move the mattress base
into the reverse-
Trendelenburg position.
If the bed is in a reverse-
Trendelenburg position,
this function will
automatically cause the
mattress base to move into
a horizontal position when
it is moved into its lowest
or highest position.
Pressing this button will
move the mattress base
into the Trendelenburg
position.
If the bed is in a
Trendelenburg position,
this function will
automatically cause the
mattress base to move into
a horizontal position when
it is moved into its lowest
or highest position.
• Note that the Trendelenburg position is only to be used on the orders
of a doctor or medical staff if the clinical condition of the patient
makes this necessary.
• Setting the Trendelenburg position unintentionally or for no reason
can endanger the resident.
• This position can only be set if the bed is connected to the mains
power supply.
Only one button can be pressed at a time, otherwise all adjustments
stop (emergency off safety function).
Reverse-
Trendelenburg