Instruction Manual for Libra Care Bed
Page 29
4.3.2 Handset (standard)
The electrical bed functions can be activated by the resident or the user by using the
handset. For safety reasons, the handset is equipped with a locking function.
Depending on the clinical condition of the patient, the user can lock handset
adjustments when deemed necessary by the supervising doctor (¼ chapter 4.3.4).
- The electric motors operate only as long
as the corresponding buttons are held
pressed.
- Adjustments are possible in both
directions.
- An elastic hook allows the handset to be
hung at practically any position on the
bed.
- The coiled cable provides ample flexibility
and freedom of movement.
- The handset can be cleaned.
- The following basic rule applies to the
buttons:
Only one button can be pressed at a time, otherwise all adjustments
stop (emergency off safety function).