4.5.3 Combined safety sides *
Please note the following regarding removed combined safety sides:
Danger of trapping and injury for the resident!
Failure to heed this warning may result in residents injuring themselves on the safety
side guides as they get out of bed if:
the upper bar of the foot-end safety side (6) has been removed and
the safety side guides are lowered.
In this case, proceed as follows in order to prevent residents from injuring themselves.
You can either:
Fit the padded cover to the centre support.
Fit the padded cover (accessory) to the lower section of the centre support. This will
cover the safety side guides and the lower section of the centre support and thus
remove the risk of injury. For details of how to install the padded cover, please refer to
the accessory manual supplied.
Or, remove the safety side guide:
Raise the foot-end safety side bars (6)
Remove the foot-end bars of the safety side
Lower the head-end safety side bars (6)
Lower the safety side guide of the safety sides (6) at the centre support completely
Remove the upper section of the centre support
Open the cover 1 and hold it firmly.
With one hand, press the release button 2 and keep it pressed.
With the other hand, take the safety side guide that faces towards the foot end 3 and
pull it straight up and out.
The resident can now get out of bed without any danger.
Reassembly: Follow the above steps in reverse order.
Attention: The safety side guide must lock into place at its end stop when inserted.
Check that it is securely fitted.