Instruction Manual for Libra Care Bed
Page 105
How should this be done?
A: Using standard handset
• Press and hold both of the marked keys
simultaneously (a rapid intermittent signal sounds)
until after approximately 5 seconds a slow intermittent
signal sounds (=RESET/ manual mode)
• Raise the mattress base
height with the “Height UP”
button to the highest horizontal
position until both motors switch off automatically
at the uppermost position, and keep the button
pressed for a further 2 seconds.
When the button is released, the intermittent
sound signal stops.
------- End of initialisation ------------
If no adjustments are possible and intermittent sound signals sound instead,
this means that the control unit has detected a fault and is locked.
➢ In this case only, carry out the following:
“RESET the control unit” (see chapter 6.5.7)