Switch mode power supply
Preparation:
o Unplug the switching power supply from the mains socket in the wall.
o Plug the switching power supply into the test socket on the test device.
Test procedure:
o Make the following selections on the instrument: Leakage current test: direct or
differential current in accordance with DIN EN 62353
o Perform a leakage current test in accordance with the instructions provided by
the test device manufacturer.
Limit value:
o Leakage current I
AB
< 0.1 mA.
Test cycles
Visual and functional check:
Yearly
Electrical measurement:
Every ten years, if the annual visual and
functional inspection was passed
We recommend the inspection cycles indicated. In the case of verifiable compliance
with the 2 % error rate (also see DGUV 3: §5, Table 1B), the inspection cycle can be
extended to a maximum of 2 years on the operator’s own responsibility.
If damage or a malfunction is suspected, the bed must be withdrawn
from service immediately and disconnected from the mains supply
until the defective parts have been repaired or replaced!
Do not make any technical changes to this product without prior
confirmation from the manufacturer!
Please use the inspection report templates included below for your inspections.