STRYKER MEDICAL/BERTEC MÉDICAL 3 - 9
MAINTENANCE GUIDE
A
B
ADJUSTMENTS
PROBLEM ENCOUNTERED
When using central motor to raise or lower bed, only the foot of bed
responds; the head does not.
Solution: adjust the slip clutch.
1. Remove the protective cover under the sleeping surface head
section. Locate the slip clutch on the Hi-Lo screw.
2. Loosen jam nut (A).
3. Use a 3/16 Allen key and tighten nut (B) one quarter of a turn.
4. Check to see how mechanism works and retighten nut (B) if
necessary.
5. Tighten jam nut (A).