3.2.4 Sample port with electrodes
The Sample port is located on the right side of the device.
The Sample port holds the Sample tube during the measurement. The start and stop
electrodes of the Sample port allow the independent determination of exactly 200 μl
volume for each sample. The method of True Volumetric Absolute Counting (TVAC)
supported by CyFlow™ Cube systems is based on the precise measurement of a fixed
sample volume.
3.2.5 Type plate
The type plate is attached to the rear side of the device.
The content of the type plate varies according to the system configuration. The system
configuration is represented by an addition to the REF number (e.g. CY-S-3061R-V2m).
Figure 5: Type plate
Table 6: Signs on the type plate of the device.