6 Operation
28
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
6.2 Performing a measurement (Sample Port)
Different measure modes are available for data acquisition. For more information, please
refer to chapter 11 Software.
Table 10: Measure Modes
Measure Mode
Description
VC electrodes
The start and stop electrode are used to measure a defined
volume of 200 µl
VC of volume
A flexible predefined volume is analyzed
Analyze all
Continuous acquisition of the complete sample (default)
Cells in Region
Continous acquisition until a flexible predefined number of
particles within in a specified region is reached
Recommendations
Make sure the correct trigger parameter is selected.
The minimal sample volume for TVAC is 850 µl.
Use flow rates between 0.5 - 5 µl/s for sample measurements. Optimal flow rates
depend on the application.
Use the same flow rate for pre-measure and measure phase to avoid fluidic
turbulences.
Requirements
The sample is prepared for measurement.
The Sheath Fluid bottle is filled.
The Waste bottle is not full.
“Sample Port” in Settings - Measure Mode - Sample Source is selected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX CyFlow Cube 6 V2m and is the answer not in the manual?

SYSMEX CyFlow Cube 6 V2m Specifications

General IconGeneral
BrandSYSMEX
ModelCyFlow Cube 6 V2m
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals