6 Operation
28
CyFlow™ Cube 6 V2m | Operating Manual | February 2020
6.2 Performing a measurement (Sample Port)
Different measure modes are available for data acquisition. For more information, please
refer to chapter 11 Software.
Table 10: Measure Modes
Measure Mode
Description
VC electrodes
The start and stop electrode are used to measure a defined
volume of 200 µl
VC of volume
A flexible predefined volume is analyzed
Analyze all
Continuous acquisition of the complete sample (default)
Cells in Region
Continous acquisition until a flexible predefined number of
particles within in a specified region is reached
Recommendations
• Make sure the correct trigger parameter is selected.
• The minimal sample volume for TVAC is 850 µl.
• Use flow rates between 0.5 - 5 µl/s for sample measurements. Optimal flow rates
depend on the application.
• Use the same flow rate for pre-measure and measure phase to avoid fluidic
turbulences.
Requirements
• The sample is prepared for measurement.
• The Sheath Fluid bottle is filled.
• The Waste bottle is not full.
• “Sample Port” in Settings - Measure Mode - Sample Source is selected.